Assessing the risk of transmission of three infectious agents among mice housed in a negatively pressurized caging system.

Document Type

Article

Publication Date

2003

Keywords

Animals-Laboratory, Bacterial-Infections, Comparative-Study, DNA-Primers, Housing-Animal, Mice, Polymerase-Chain-Reaction

First Page

16

Last Page

21

JAX Location

see Journal Stacks

JAX Source

Contemp Top Lab Anim Sci 2003 Nov; 42(6):16-21.

Abstract

Previous studies from our institution have shown that ventilated caging run at negative pressure to a mouse room dramatically reduced exposure of personnel to the major mouse allergen, Mus m 1. The current study was designed to determine whether negative cage ventilation posed an inordinate risk for spread of infectious agents between cages and/or racks. B6;129S-Tnfsf5(tm1Imx)/J (TNF) mice, which were naturally and persistently infected with Pneumocystis carinii, Helicobacter bilis, and Pasteurella pneumotropica, were used as the source of infections. Uninfected C3Smn.CB17-Prkdc(scid)/J (SCID) mice with severe combined immunodeficiency were used to detect transmission. The following methods were used to detect transmission of infections: polymerase chain reaction (PCR) amplification and histological examination of lungs for P. carinii; PCR of fecal specimens or cecal contents for H. bilis; and culture of oropharyngeal, tracheal, or vaginal swabs for P. pneumotropica. We determined whether transmission of the three agents occurred via direct contact (cohabitation), exposure to soiled bedding, and/or by handling naive SCID mice after handling infected TNF mice. During a 12-week period, all three infectious agents were readily transmitted to uninfected mice by cohabitation. Transmission was much less efficient and occurred later among mice exposed to contaminated bedding. Transmission did not occur as a result of handling. We then studied transmission of the three infectious agents among mice housed in individually ventilated cages run at negative pressure in a small, crowded mouse room. Transmission of P. carinii was detected at the end of the 12-month study in the densely populated room, probably because the exhaust from the changing station passed over soiled cages and caused aerosolization of particulates. Caging systems run at negative pressure effectively reduce personnel exposure to allergens and may also inhibit the transmission of infectious diseases.

Please contact the Joan Staats Library for information regarding this document.

Share

COinS